Without touching the alignment in to the diagnostic breadboard, a scan of the front end laser was done;
partly to see what the status of the diagnostic breadboard is after being dormant for such a long time.
The mode scan implies that the output of the front end has 61% in higher order modes. Something we
know not to be true. So clearly some alignment work into the diagnostic breadboard is required (this
is not a surprise).
It should also be noted that whilst the scan was conducted the PMC PZT monitor flashed red. But the
take away news is that the diagnostic breadboard still works but needs a service call.