Laser Status:
SysStat is good
Front End Power is 35.95W (should be around 30 W)
HPO Output Power is 153.1W
Front End Watch is GREEN
HPO Watch is GREEN
PMC:
It has been locked 0 days, 11 hr 8 minutes (should be days/weeks)
Reflected power = 20.4Watts
Transmitted power = 49.1Watts
PowerSum = 69.5Watts.
FSS:
It has been locked for 0 days 11 hr and 7 min (should be days/weeks)
TPD[V] = 3.184V (min 0.9V)
ISS:
The diffracted power is around 3.6% (should be 3-5%)
Last saturation event was 5 days 1 hours and 55 minutes ago (should be days/weeks)
Possible Issues:
PMC reflected power is high