Displaying report 1-1 of 1.
Reports until 09:31, Tuesday 03 October 2017
H1 PSL
jason.oberling@LIGO.ORG - posted 09:31, Tuesday 03 October 2017 (38865)
PSL Weekly FAMIS Tasks (FAMIS 3670 & 8442)

This morning I completed the weekly PSL FAMIS tasks.

HPO Pump Diode Current Adjust (FAMIS 8442)

With the ISS OFF, all operating currents of the HPO pump diode boxes were increased by 0.1 A; the changes are summarized in the below table.  A screenshot of the PSL Beckhoff main screen is attached for future reference.

  Operating Current (A)
Old New
DB1 50.4 50.5
DB2 52.9 53.0
DB3 52.9 53.0
DB4 52.9 53.0

The operating temperatures of the diode boxes were not adjusted.  With the ISS back ON, the HPO is outputting 153.7 W.  This completes FAMIS 8442.

PSL Power Watchdog Reset (FAMIS 3670)

I reset both PSL power watchdogs at 16:19 UTC (9:19 PDT).  This completes FAMIS 3670.

Images attached to this report
Displaying report 1-1 of 1.