H1 TCS
thomas.shaffer@LIGO.ORG - posted 17:12, Monday 09 October 2017 - last comment - 17:23, Tuesday 17 October 2017(38956)
Installed SR3 RoC Heater

Betsy, Hugh, TJ

Last week Betsy put the heater on the table and today Hugh checked it's vertical center with an auto-level. Some washers were added to get it as close to center as possible, which ended up around 0.4mm high. Betsy and I then had to wiggle the assembly into place, and rotate the entire gold ceramic holder to allow the screws on the outside to clear the OSEM brackets. The heater is currently sitting ~6mm away from the back of the SR3 optic and it is plugged into the feed through.

Picture attached.

Images attached to this report
Comments related to this report
betsy.weaver@LIGO.ORG - 17:24, Monday 09 October 2017 (38957)

Here are a few more pics.  As TJ notes, the ROC front face is 5-6mm from the SR3 AR surface.  It is locked down in this location. 

 

Note, we followed a few hints from LLO's install:

https://alog.ligo-la.caltech.edu/aLOG/index.php?callRep=25831

 

Continuity checks at the feedthru still need to be made.  Will solicit EE for their help.

Images attached to this comment
filiberto.clara@LIGO.ORG - 12:42, Tuesday 10 October 2017 (38969)

Initial continuity test failed. Found issues with in-vacuum cable, power pins not pushed in completely. Pins were pushed in until a locking click was heard.

Reading are:
Larger Outer Pins, Heater: 66.9Ω
Inner pair (left most looking at connector from air side), thermacouple: 105Ω

hugh.radkins@LIGO.ORG - 15:24, Tuesday 10 October 2017 (38974)

Found center of SR3(-X Scribe) to be 230.2mm above optical table.  By siting the top and bottom of RoC Heater, found center to be at 231.4mm. Removed available shim to put center of RoC Heater at 229.6 for 230.2-229.6=0.6mm below perfect.

thomas.shaffer@LIGO.ORG - 08:27, Wednesday 11 October 2017 (38981)

Hugh's comment reminded me that to get the heater to fit, Betsy and I added a 1mm washer to raise the height of the assembly. In total we have 4mm of washers (2x1.5mm & 1x1mm).

stephen.appert@LIGO.ORG - 17:23, Tuesday 17 October 2017 (39058)TCS
I conducted measurement of quantity 6 of [D1600104 SR3 ROC Actuator, Ceramic Heater Assy] at CIT on 4 March 2016. Dirty state before baking.
The serial number of the heater assy installed in LHO HAM5 is S1600180 - see https://ics-redux.ligo-la.caltech.edu/JIRA/browse/ASSY-D1500258-002
S1600180 Resistance = 66.8 Ohms on 4 March 2016.
There is good agreement between the as-installed and pre-bake measurements.