There was some ambiguity as to whether the NPRO noise eater was on or off because the MEDM screen
did not change.
    Attached is a plot of the NPRO output as seen by the photodetector inside the front end laser
with the noise eater toggled on and off.  Clearly the switch on the power supply works.  However
the status indicator did not change.  The output voltage of the photodetector was 0.210 V and the
number of counts on the MEDM screen was ~ -335.  Blocking the light to the photodiode resulted in
the status indicator changing and an increase in the number of counts to ~ -5000.  We know that
the change in status occurs around ~ -1792 counts.
    I do not recall when this problem first manifested itself.