Laser Status:
SysStat is good
Front End Power is 35.93W (should be around 30 W)
HPO Output Power is 154.0W
Front End Watch is GREEN
HPO Watch is GREEN
PMC:
It has been locked 2 days, 18 hr 41 minutes (should be days/weeks)
Reflected power = 23.4Watts
Transmitted power = 48.02Watts
PowerSum = 71.42Watts.
FSS:
It has been locked for 0 days 13 hr and 45 min (should be days/weeks)
TPD[V] = 2.632V (min 0.9V)
ISS:
The diffracted power is around 3.0% (should be 3-5%)
Last saturation event was 2 days 18 hours and 40 minutes ago (should be days/weeks)
Possible Issues:
PMC reflected power is high