FAMIS 7461 Laser Status: SysStat is good Front End Power is 35.89W (should be around 30 W) HPO Output Power is 153.9W Front End Watch is GREEN HPO Watch is GREEN PMC: It has been locked 3 days, 19 hr 37 minutes (should be days/weeks) Reflected power = 22.64Watts Transmitted power = 47.78Watts PowerSum = 70.43Watts. FSS: It has been locked for 0 days 16 hr and 19 min (should be days/weeks) TPD[V] = 2.789V (min 0.9V) ISS: The diffracted power is around 3.3% (should be 3-5%) Last saturation event was 3 days 9 hours and 1 minutes ago (should be days/weeks) Possible Issues: PMC reflected power is high