J. Kissel Now that MC2 has been fixed (LHO aLOG 39382), other than PR3 (LHO aLOG 39383), all intentionally freely suspended HAM2 / HAM3 suspensions are free of rubbing at the moment: MC1, MC2, MC3, IM1, IM2, IM3, IM4, PRM, PR3. Note, PR2 still has its protective lens cap on, and likely EQ stopped so I didn't both measuring it. Data can be found commited with the 2017-11-10 date stamp in the appropriate folder of the SusSVN. /ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/MC?/SAGM1/Data/*.xml /ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/PRM/SAGM1/Data/*.xml /ligo/svncommon/SusSVN/sus/trunk/HLTS/H1/PR3/SAGM1/Data/*.xml /ligo/svncommon/SusSVN/sus/trunk/HAUX/H1/IM?/SAGM1/Data/*.xml