Displaying report 1-1 of 1.
Reports until 09:25, Monday 13 November 2017
H1 PSL
thomas.shaffer@LIGO.ORG - posted 09:25, Monday 13 November 2017 - last comment - 09:27, Monday 13 November 2017(39403)
PSL Weekly Report

Laser Status:
SysStat is good
Front End Power is 35.82W (should be around 30 W)
HPO Output Power is 153.0W
Front End Watch is GREEN
HPO Watch is GREEN

PMC:
It has been locked 0 days, 0 hr 2 minutes (should be days/weeks)
Reflected power = 22.43Watts
Transmitted power = 45.97Watts
PowerSum = 68.4Watts.

FSS:
It has been locked for 0 days 0 hr and 2 min (should be days/weeks)
TPD[V] = 1.931V (min 0.9V)

ISS:
The diffracted power is around 4.2% (should be 3-5%)
Last saturation event was 0 days 0 hours and 1 minutes ago (should be days/weeks)

Possible Issues:
PMC reflected power is high

Comments related to this report
thomas.shaffer@LIGO.ORG - 09:27, Monday 13 November 2017 (39404)

FAMIS7464 closed.

Displaying report 1-1 of 1.