Repeated isolation measurement of the OFI in the optics lab

[Gerardo, Koji]

The isolation measurement of the OFI was repeated today to confirm the stability of the optical isolation performance. For the confirmation, we have used the usual photodiode technique, as well as the power meter. Also we made the dry run of the in-situ isolation test setup [LHO ALOG 39357]. All the measurements showed the consistent isolation better than 40dB. We are going to install the OFI optical table to HAM5 tomorrow.

Methods

Results

  LHO OFI 2017/11/13
After the adjustment
[LHO ALOG 39418]
LHO OFI 2017/11/14
PD measurement

 
LHO OFI 2017/11/14
Power meter measurement
LHO OFI 2017/11/14
In-situ measurement
Temperature of
the faraday rotator body
N/A N/A 73degF = 22.8degC 76degF = 24.4degC
Transmission
From the input port to the output port
0.967 +/- 0.002
(3.3% loss)
--- --- ---
Back-scatter isolation S-pol
From the input port to the squeezer port
1300 +/- 10 ppm --- --- ---
TFP AR reflection
From the input port to the squeezer port
890 +/- 10 ppm --- --- ---
Isolation
From the output port to the input port
52 +/- 13 ppm
( = 43 +/- 1 dB isolation)
48.4 +/- 0.7 ppm
( = 43.15 +/- 0.05 dB isolation)
42 +/- 2 ppm
( = 43.7 +/- 0.3 dB isolation)
34.5 +/- 0.4 ppm
( = 44.6 +/- 0.5 dB isolation)
Back-scattering
From the input port to the input port
  3.0 +/- 0.5 ppm 0.5 +/- 0.2 ppm 2 +/- 3 ppm ---
Squeezer transmission
From the squeezer port to the input port
0.978 +/- 0.002
(2.2% loss)
--- --- ---