[Gerardo, Koji]
The isolation measurement of the OFI was repeated today to confirm the stability of the optical isolation performance. For the confirmation, we have used the usual photodiode technique, as well as the power meter. Also we made the dry run of the in-situ isolation test setup [LHO ALOG 39357]. All the measurements showed the consistent isolation better than 40dB. We are going to install the OFI optical table to HAM5 tomorrow.
Methods
Results
LHO OFI 2017/11/13 After the adjustment [LHO ALOG 39418] |
LHO OFI 2017/11/14 PD measurement |
LHO OFI 2017/11/14 Power meter measurement |
LHO OFI 2017/11/14 In-situ measurement |
|
Temperature of the faraday rotator body |
N/A | N/A | 73degF = 22.8degC | 76degF = 24.4degC |
Transmission From the input port to the output port |
0.967 +/- 0.002 (3.3% loss) |
--- | --- | --- |
Back-scatter isolation S-pol From the input port to the squeezer port |
1300 +/- 10 ppm | --- | --- | --- |
TFP AR reflection From the input port to the squeezer port |
890 +/- 10 ppm | --- | --- | --- |
Isolation From the output port to the input port |
52 +/- 13 ppm ( = 43 +/- 1 dB isolation) |
48.4 +/- 0.7 ppm ( = 43.15 +/- 0.05 dB isolation) |
42 +/- 2 ppm ( = 43.7 +/- 0.3 dB isolation) |
34.5 +/- 0.4 ppm ( = 44.6 +/- 0.5 dB isolation) |
Back-scattering From the input port to the input port |
3.0 +/- 0.5 ppm | 0.5 +/- 0.2 ppm | 2 +/- 3 ppm | --- |
Squeezer transmission From the squeezer port to the input port |
0.978 +/- 0.002 (2.2% loss) |
--- | --- | --- |