PSL Weekly Status Report


Laser Status:
SysStat is good
Front End Power is 35.76W (should be around 30 W)
HPO Output Power is 153.0W
Front End Watch is GREEN
HPO Watch is GREEN

PMC:
It has been locked 9 days, 8 hr 47 minutes (should be days/weeks)
Reflected power = 23.77Watts
Transmitted power = 48.68Watts
PowerSum = 72.44Watts.

FSS:
It has been locked for 2 days 3 hr and 28 min (should be days/weeks)
TPD[V] = 2.294V (min 0.9V)

ISS:
The diffracted power is around 2.3% (should be 3-5%)
Last saturation event was 6 days 6 hours and 12 minutes ago (should be days/weeks)

Possible Issues:
PMC reflected power is high

 