Ran the checks for the STS & T240 masses. STS's were fine. The T240 had the following masses out of range:
Averaging Mass Centering channels for 10 [sec] ...
2017-12-04 10:39:03.361987
There are 20 T240 proof masses out of range ( > 0.3 [V] )!
ETMX T240 1 DOF X/U = 0.335 [V]
ETMX T240 2 DOF X/U = -0.379 [V]
ETMX T240 2 DOF Y/V = -0.709 [V]
ETMX T240 3 DOF Y/V = 0.337 [V]
ETMY T240 1 DOF Y/V = 0.323 [V]
ETMY T240 3 DOF Z/W = 0.364 [V]
ITMX T240 1 DOF X/U = -0.387 [V]
ITMX T240 2 DOF X/U = -0.32 [V]
ITMX T240 3 DOF X/U = -0.628 [V]
ITMY T240 1 DOF X/U = -0.338 [V]
ITMY T240 1 DOF Z/W = -0.538 [V]
ITMY T240 2 DOF Z/W = -0.609 [V]
ITMY T240 3 DOF X/U = -1.026 [V]
ITMY T240 3 DOF Y/V = -0.341 [V]
ITMY T240 3 DOF Z/W = -0.759 [V]
BS T240 1 DOF Y/V = -0.35 [V]
BS T240 1 DOF Z/W = -0.519 [V]
BS T240 2 DOF Y/V = -0.443 [V]
BS T240 3 DOF X/U = -0.637 [V]
BS T240 3 DOF Z/W = -0.564 [V]
2017-12-04 10:37:23.352027
All STSs proof masses that within healthy range (< 2.0 [V]). Great!
Closing out this FAMIS task.