Reports until 10:44, Tuesday 05 December 2017
H1 PSL
jason.oberling@LIGO.ORG - posted 10:44, Tuesday 05 December 2017 (39635)
PSL Weekly FAMIS Tasks (FAMIS 4022 & 8451)

This morning I completed the weekly PSL FAMIS tasks.

HPO Pump Diode Current Adjust (FAMIS 8451)

With the ISS OFF, I adjusted the HPO pump diode operating currents.  The changes are summarized in the below table; a screenshot of the PSL Beckhoff main screen is attached for future reference.

  Operating Current (A)
Old New
DB1 53.0 53.2
DB2 54.8 55.0
DB3 54.8 55.0
DB4 54.8 55.0

I did not adjust the operating temperatures of the diode boxes.  The HPO is now outputting 155.4 W and the ISS is back ON.  This completes FAMIS 8451.

PSL Power Watchdog Reset (FAMIS 4022)

I reset both PSL power watchdogs at 18:35 UTC (10:35 PST).  This completes FAMIS 4022.

Images attached to this report