Reports until 11:17, Monday 18 December 2017
H1 PSL
thomas.shaffer@LIGO.ORG - posted 11:17, Monday 18 December 2017 (39796)
PSL Weekly Report

FAMIS7469


Laser Status:
SysStat is good
Front End Power is 35.66W (should be around 30 W)
HPO Output Power is 152.9W
Front End Watch is GREEN
HPO Watch is GREEN

PMC:
It has been locked 6 days, 3 hr 58 minutes (should be days/weeks)
Reflected power = 25.6Watts
Transmitted power = 48.03Watts
PowerSum = 73.64Watts.

FSS:
It has been locked for 5 days 0 hr and 29 min (should be days/weeks)
TPD[V] = 2.725V (min 0.9V)

ISS:
The diffracted power is around 1.8% (should be 3-5%)
Last saturation event was 6 days 2 hours and 14 minutes ago (should be days/weeks)

Possible Issues:
PMC reflected power is high
ISS diffracted power is Low