Reports until 11:04, Tuesday 19 December 2017
H1 PSL
jason.oberling@LIGO.ORG - posted 11:04, Tuesday 19 December 2017 (39818)
PSL Weekly FAMIS Tasks (FAMIS 4024 & 8453)

This morning I completed the weekly PSL FAMIS tasks.

HPO Pump Diode Operating Current Adjust (FAMIS 8453)

With the ISS OFF, I adjusted the operating current of the HPO DBs.  The changes are summarized in the below table and a screenshot of the main PSL Beckhoff screen is attached for future reference.

  Operating Current (A)
Old New
DB1 53.4 53.7
DB2 55.2 55.4
DB3 55.2 55.4
DB4 55.2 55.4

I did not adjust the operating temperatures of the HPO DBs.  The HPO is now outputting 155.2 W and the ISS is back ON.  This completes FAMIS 8453.

PSL Power Watchdog Reset (FAMIS 4024)

I reset both PSL power watchdogs at 18:54 UTC (10:54 PST).  This completes FAMIS 4024.

Images attached to this report