Displaying report 1-1 of 1.
Reports until 08:26, Wednesday 15 November 2017
H1 CAL (CAL)
madeline.wade@LIGO.ORG - posted 08:26, Wednesday 15 November 2017 - last comment - 06:46, Wednesday 20 December 2017(39419)
Filters to use for C02 frame generation

[Maddie W., Aaron V., Greg M.]

Here is a summary of the filters and command line arguments that should be used for the C02 frame generation using the DCS calibration pipeline.  This information is compiled from configurations wiki page and aLOGs linked from that wiki page.  The command line arguments should be the same as for C01 except for changes noted in the table below.  (Additionally, the frame names and channel names should be changed appropriately to C02.)

GPS times Filters files Special command line arguments Change from previous epoch

1163173888 - 1167436818

aligocalibration/trunk/Runs/O2/GDSFilters/H1DCS_1163173888.npz
--expected-fcc=347.2
--coherence-uncertainty-threshold=0.004
--wings=(something larger than 600)
--no-fs 
--no-srcQ
--update-fcc
--fcc-averaging-time=600
--fcc-filter-taper-length=32768
--pcal-channel-name CAL-PCALY_TX_PD_OUT_DQ
N/A

1167436818 - 1169326080

aligocalibration/trunk/Runs/O2/GDSFilters/H1DCS_1167436818.npz
--expected-fcc=360.0
--coherence-uncertainty-threshold=0.004
--wings=(something larger than 600)
--no-fs
--no-srcQ
--update-fcc
--fcc-averaging-time=600
--fcc-filter-taper-length=32768
--pcal-channel-name CAL-PCALY_TX_PD_OUT_DQ

Calibration model changes

New H1 params file created on 1/3/2017

Coupled cavity pole frequency changed in reference model

1169326080 - 1173225472

aligocalibration/trunk/Runs/O2/GDSFilters/H1DCS_1169326080.npz same as above New H1 params file created on 1/24/2017 (see aLOG 33585)

1173225472 - end of run

aligocalibration/trunk/Runs/O2/GDSFilters/H1DCS_1173225472.npz
--expected-fcc=360.0
--coherence-uncertainty-threshold=0.004 
--wings=(something larger than 600)
--update-fcc
--fcc-averaging-time=600
--fcc-filter-taper-length=32768
--pcal-channel-name CAL-PCALY_TX_PD_OUT_DQ

Include EPICS to compute SRC detuning parameters

NOTE: Starting with these frames there will be additional channels in the frames for the SRC detuning parameters

Comments related to this report
madeline.wade@LIGO.ORG - 19:58, Monday 20 November 2017 (39492)

After a discussion with the review team, we are just going to use a taper length of 32768 for both LHO and LLO.  I have therefore just made this the default in the code.  You do not need to set the --fcc-filter-taper-length option for C02 frame generation.

madeline.wade@LIGO.ORG - 06:46, Wednesday 20 December 2017 (39832)

We have finished our rounds of testing and are settling on the options used to produce the X05 test frames (summary page here). Below is an updated table of the command lines to be used:

 

GPS times Filters files Special command line arguments Change from previous epoch

1163173888 - 1167436818

aligocalibration/trunk/Runs/O2/GDSFilters/H1DCS_1163173888.npz
--expected-fcc=347.2
--coherence-uncertainty-threshold=0.02
--wings=(something larger than 300)
--no-fs 
--no-srcQ
--update-fcc
--fcc-averaging-time=60
--pcal-channel-name CAL-PCALY_TX_PD_OUT_DQ
n/a

1167436818 - 1169326080

aligocalibration/trunk/Runs/O2/GDSFilters/H1DCS_1167436818.npz
--expected-fcc=360.0
--coherence-uncertainty-threshold=0.02
--wings=(something larger than 300)
--no-fs
--no-srcQ
--update-fcc
--fcc-averaging-time=60
--pcal-channel-name CAL-PCALY_TX_PD_OUT_DQ

Calibration model changes

New H1 params file created on 1/3/2017

Coupled cavity pole frequency changed in reference model

1169326080 - 1173225472

aligocalibration/trunk/Runs/O2/GDSFilters/H1DCS_1169326080.npz same as above New H1 params file created on 1/24/2017 (see aLOG 33585)

1173225472 - end of run

aligocalibration/trunk/Runs/O2/GDSFilters/H1DCS_1173225472.npz
--expected-fcc=360.0
--coherence-uncertainty-threshold=0.02 
--wings=(something larger than 300)
--update-fcc
--fcc-averaging-time=60
--pcal-channel-name CAL-PCALY_TX_PD_OUT_DQ

Include EPICS to compute SRC detuning parameters

NOTE: Starting with these frames there will be additional channels in the frames for the SRC detuning parameters

Displaying report 1-1 of 1.