kyle.ryan@LIGO.ORG - posted 18:43, Thursday 21 December 2017 (39884)
Misc. Vacuum stuff
Mark D., Tyler G. and Bubba G. having tacked on both of HAM3's doors this afternoon has left us with only three chamber door openings with soft-covers on.  As such, we had to dial back the purge air supply from its recent nominal value of 100% open to something much less so as to not blow these three covers off!  Again, not yet having a flow meter in-line with the purge-air supply results in many theories etc.. but everything now seems typical of the historical behavior with the purge air.  Also, the dew point is dropping as a result - two hours after HAM3's doors went on it has dropped down to -31C from a months long steady value of around -28C.  

BSC1's annulus pressure is still good and getting better (see discussion from yesterday).  BSC3's annulus pressure, on the other hand, had seemed to have bottomed out at the pump cart (2.1 x 10-5 Torr) since yesterday.  So, this evening, I closed the annulus isolation valve and noted that the pump cart pressure didn't respond.  Having a leak on the pump side of the isolation valve "muddies the water" a little bit.  We don't have the luxury of starting the ion pump and monitoring the current over a period of days - so, we will try and spray a little helium tomorrow so as to rule out an outer O-ring leak that might be getting "masked" my the pump-side leak.  

The fourth of 4 NEG pumps is finishing up a "cycle" of particulate mitigation in the lab and should make the deadline tomorrow to be available to get installed.