[Kyle, Chandra]
Started both QDP80 roughing pumps this morning with a purge air load to let them warm up before valving into corner volume. Blocks measure ~ 70degC.
Kyle checked the pressure of main volume to verify we aren't over pressurizing system with purge air - measured 1/2 psi with calibrated mechanical gauge that we use for proof testing viewports. PT-120A pirani gauge on BSC2 reads 942 Torr, which converts to 3.5 psi over standard atmospheric pressure, but we know this gauge is biased high.