kyle.ryan@LIGO.ORG - posted 15:09, Saturday 30 December 2017 - last comment - 17:05, Saturday 30 December 2017(39938)
1505 hrs. local -> Kyle on site checking misc. vacuum pumps etc..
I'll go ahead and do an impromptu site inspection while I'm here. I expect to be here between 60- 90 minutes.
I will make a comment to this log entry when I leave
Comments related to this report
kyle.ryan@LIGO.ORG - 17:05, Saturday 30 December 2017 (39939)
Vacuum pumps and purge air are all normal.
local gauge indications Vertex MTP (7.7 x 10-7 torr, 1.0 x 10-2 torr), XBM MTP (8.8 x 10-7 torr, 6.9 x 10-3 torr) and YBM MTP (1.1 x 10-6 torr, 8.5 x 10-3 torr)
SITE INSPECTION:
1) JOY fan (HVAC) at X-mid is quite whiney/screatchy and noticeablly noiser than the rest of the fans on site -> I was unable to open the door to inspect as the internals of the door handle seem broken.
2) Filler cap on main chiller (ante room?) was not in place, i.e. just dangling. -> I left as found.
3) Weld shop door was found to be unlocked -> I locked it.
4) VPW above-ground waste water tank must be full as the "RED" lamp is illuminated.
1705 hrs. local -> Kyle leaving site now, no one else here at this time.