Richard informed me that the laser had tripped this morning. I reset the laser without any problems.
I increased the current for head 1 to the maximum of 60A and increased the diode temperature for
diode 3 to 23 degC. The output of diode box 1 has taken a bit of a dive since the weekend. So much
so that I would start to become concerned that the thermal lens in the NdYAG rod isn't what it should
be to make the resonator stable.
At this point in time, we have two options as I see it, depending on commissioning constraints:
i. replace the diode box
ii. revert to low power mode
Both would take about half a day to do.
Attached is a trend of the DB operating current increases, starting in April 2017. The ongoing failure of DB1 can be seen in the large operating current increases that have been required over the last 2 weeks to keep the laser running. The life of DB1 has been slightly extended by adjusting its operating temperature in lieu of adjusting its operating current, hence the flat spot near the end of the graph.
If your trying to stretch out the lifetime I seem to recall that you can actually run these diodes up to 65A in a pinch