Laser Status:
SysStat is good
Front End Power is 34.81W (should be around 30 W)
HPO Output Power is 29.17W
Front End Watch is GREEN
HPO Watch is RED
PMC:
It has been locked 1 days, 15 hr 1 minutes (should be days/weeks)
Reflected power = 3.255Watts
Transmitted power = 22.57Watts
PowerSum = 25.82Watts.
FSS:
It has been locked for 0 days 2 hr and 9 min (should be days/weeks)
TPD[V] = 0.7296V (min 0.9V)
ISS:
The diffracted power is around 2.7% (should be 3-5%)
Last saturation event was 1 days 15 hours and 2 minutes ago (should be days/weeks)
Possible Issues:
FSS TPD is low