J. Kissel, B. Weaver Far more details to come, but we've replaced the H1 SUS ITMY R0 LF OSEM in order to fix the noise / shorting problem (see LHO aLOG 40555, FRS Ticket 9683, Work Permit 7358). It's new open light current values are Open Light Current / ct GAIN OFFSET (as ADC Counts) (30000/olc) (-olc/2) R0 RT 29804 1.007 -14902 I've accepted the new values in the SDF system.
Serial number 466 was removed and serial number 298 was installed.