H1 PSL (PSL)
matthew.heintze@LIGO.ORG - posted 17:09, Thursday 22 February 2018 - last comment - 10:07, Friday 23 February 2018(40673)
70W progress Thurs 22

Maik, Jason, Matt

 

Semi frustrating day today. However we have:

 

We also could not test the mechanism for the new external shutter, as the mechanism is currently just bare wires and so we need to get the pinout to wire up a connector (this wasnt the case at LLO as it already had the connectors wires)

And we are having the issues with the diode chiller that we still havent been able to rectify.

Comments related to this report
jeffrey.bartlett@LIGO.ORG - 10:07, Friday 23 February 2018 (40683)
   The new diode chiller could be run from local control with no apparent problems. However, it would not start from Beckhoff. It appeared to Beckhoff that the chiller was still on. Several things were tried but could not get the chiller to start from Beckhoff. 

    This morning, the diode chiller controller from the old chiller(Gromit)was installed in the new chiller (Wallace). On power up, the diode chiller started without command input, which it should not do. We stopped and started both chiller from Beckhoff several times and both appear to be working correctly.

  I will contact the manufacture of the faulty controller for repair or replacement.    