This morning, Kyle and Bubba finished removing the door to BSC1. I entered the chamber and began poking the ITMY R0 Right BOSEM cable from the OSEM connection out through the structure to the Table Cable Bracket. No change. I then also tried all of the following which produced no improvement to the spectra:
- Reseating the cable in the Table Cable Bracket
- Unbundling the cables that were secured to the upper structure leg
- Reseating the cable on the BOSEM itself
- Swapped the RT cable with the Side - problem went away - which meant that the problem was the BOSEM itself.
Removed BOSEM from structure and attempted to find short on the unit - no luck. Swapped unit which meant removing it from it's adjustment bracket and rebuilding before reattaching.
Once back on the structure found it was OK. Set OLV and 50% flag setting.
Cleaned flooring, removed tools, looked for cable routing issues in case I undid some other nice work.
Left chamber. Kissel ran a few TFs which showed healthy. Fil checked for HIPOT on the ESD connection and started another rounf of ground loop checks. Tacked up door.