Displaying report 1-1 of 1.
Reports until 10:02, Thursday 01 March 2018
H1 SUS
jeffrey.kissel@LIGO.ORG - posted 10:02, Thursday 01 March 2018 (40787)
H1SUSITMY Remains Clear of Electrical Short-to-Ground Symptoms After Pump-Down
J. Kissel
FRS Ticket 9683

Checking in on the top-mass health of H1 SUS ITMY after the corner station volume has passed below 1e-5 Torr, I've gathered an high frequency, amplitude spectral density of the Main (M0) and Reaction (R0) chain top mass OSEMs. All sensors show a flat noise floor consistent with the expected self noise of the OSEM (~5e-11 / m.Hz^{-1/2} above 10 Hz)  and no longer show any significant comb-like features. 

As such, I recommend we close out the above cited FRS Ticket.
Images attached to this report
Displaying report 1-1 of 1.