Reports until 13:17, Wednesday 07 March 2018
H1 SQZ
daniel.sigg@LIGO.ORG - posted 13:17, Wednesday 07 March 2018 - last comment - 16:31, Wednesday 07 March 2018(40881)
VOPO Work

TJ Sheila Daniel

TJ fixed the short on DCPD2 betwen anode and case following Rich's suggestion of adding a piece of viton under the ceramic circuit board. The same was done for DCPD1.

However, wew found that for DCPD1 the anode and cathode are swapped. We made it work with some clip leads (for now).

We tested the PZTs and both of them are working.

Comments related to this report
thomas.shaffer@LIGO.ORG - 16:31, Wednesday 07 March 2018 (40884)

Small correction, I added a piece of Kapton tubing that we normally use to help route/control cables. I rolled the tubing and then cut a small slit vertically so that I could stick the top pin through and hold it in place. Picture attached.

Images attached to this comment