kyle.ryan@LIGO.ORG - posted 07:43, Sunday 11 March 2018 - last comment - 02:39, Monday 12 March 2018(40950)
CP4 bake-out controller in "HOLD" again
Output has been at 100& for days now so this won't impact anything.  If situation were different, i.e. were weren't already at 100% output but still ramping up, then this "bug" could have resulted in an over temperature condition.  The program is behaving as if the "HOLD" condition is a function of run time based on 1C / hour expected rate of rise and time needed to get to SETPOINT value from measured temperature recorded at start of profile.  If so, then this is a novice programmer mistake.  The program should switch to the HOLD segment upon reaching the SETPOINT value as a function of the measured temperature irregardless of time so as to always be limited to the 1C / hour maximum rate of rise.   
Comments related to this report
chandra.romel@LIGO.ORG - 08:30, Sunday 11 March 2018 (40951)

I also witnessed this happen on Friday when the control sensor clearly hadn't reached the set point yet and program transitioned to hold at temp. Heater had been at 100% output overnight.

gerardo.moreno@LIGO.ORG - 02:39, Monday 12 March 2018 (40952)VE

Temperature at GV11 in air for the past 3 days.