Displaying report 1-1 of 1.
Reports until 08:20, Monday 12 March 2018
H1 General (PSL)
edmond.merilh@LIGO.ORG - posted 08:20, Monday 12 March 2018 (40956)
PSL Status Report - Weekly: FAMIS #7481


Laser Status:
SysStat is good
Front End Power is 33.97W (should be around 30 W)
HPO Output Power is -0.04668W
Front End Watch is GREEN
HPO Watch is RED

PMC:
It has been locked 0 days, 0 hr 0 minutes (should be days/weeks)
Reflected power = 0.02522Watts
Transmitted power = -0.02523Watts
PowerSum = -0.001669Watts.

FSS:
It has been locked for 0 days 0 hr and 0 min (should be days/weeks)
TPD[V] = 0.0672V (min 0.9V)

ISS:
The diffracted power is around 3.3% (should be 3-5%)
Last saturation event was 6 days 19 hours and 17 minutes ago (should be days/weeks)

Possible Issues:
FSS TPD is low
LRA out of range, see SYSSTAT.adl

 