TCSY glitching signals - an update to previous alog 40843
In my alog 40843, the drawing I included shows the laser thermistor being read into box D1300649, which is incorrect. The laser temperature is an un-labled connection further down the drawing that shows the signal going into the control box, D1200745.
That said, both boxes have signals that either glitch or drop in DC level at the time that the flow rate glitches.
The instant recovery of the flow rate doesn't match an actual recovery due to a temporary reduction in flow due to debris in the chiller lines.
With Fil's help, I can confirm that the D1300649 box, and the D1200745 box, and the flow meter have all been swapped out, and the glitches persist.
Attached files show the glitching signals and the where they're read.
Images attached to this report