Reports until 17:24, Monday 12 March 2018
H1 SQZ
daniel.sigg@LIGO.ORG - posted 17:24, Monday 12 March 2018 (40976)
Squeezer TTFSS

Terry Daniel

The TTFSS slow output is busted again. We discovered the same symptom with the temperature control input having a -14V offset. This seems to put the laser in a funky state: operating multi-mode would explain some of the other strange things we encountered before we identified the problem. At the end of the day, we were back were we left on Friday...