Reports until 18:39, Wednesday 14 March 2018
H1 SQZ
daniel.sigg@LIGO.ORG - posted 18:39, Wednesday 14 March 2018 (41017)
OPO locked in-air

Nutsinee Terry Daniel

We were finally able to lock the OPO in air. We took the output of the common mode board and send it to the laser PZT via a Thorlabs PZT driver (gain of 15).  The ugf we achieved is 10kHz. With all 3 boosts engaged the lock is reasonably tight. The attached screenshot shows the CMB settings.

The noise spectrum indicates a strong acoustic peak around 1.1kHz that needs a large driving range.

We also measured the transfer function from the OPO PZT to the laser PZT. The first resonance is at ~6.0 kHz as expected—confirming that the 1.1kHz peak is not an inline resonance.

Some of the problems we encountered along the way:

Images attached to this report