- original glitch, shown in alog 40965, now showing both TCSY and TCSX
- this shows that where the TCSY laser temperature, chiller flow, and QPD B seg4 all glitched, nothing is seen on TCSX signals
- all 3 signals from TCSY glitch independently of TCSX
- new glitch
- both TCSY and TCSX QPD B seg4's gltich, but no other signals do
- TCSY and TCSX QPDs glitch at the same time
- new glitch
- both TCSY and TCSX QPD B seg4's gltich, and at the same time both laser temperatures start a slow rise in value, right before TCSX laser is turned on
- this variety may have been due to preparations for turning the TCSX laser on, however, if that's true, it's significant that both TCSX and TCSY signals were effected
- TCSY and TCSX QPDs and laser temperatures glitch at the same time
From these plots, I would say that it's very possible that there are multiple sources of glitching in the TCS system.