Displaying report 1-1 of 1.
Reports until 13:22, Monday 26 March 2018
H1 SQZ (SQZ)
sheila.dwyer@LIGO.ORG - posted 13:22, Monday 26 March 2018 - last comment - 14:01, Wednesday 28 March 2018(41150)
OPO phase matching in air

On Friday Terry and I adjusted the crystal position in the OPO to get phase matching and co-resonance at the same temperature. The take away messages are that the crystal translation stage is working and if we correct for mode matching the OPO threshold is between 7.4-8mW, and our phases matching temperature setting in air is just below 34C, both of these are consistent with our expectations.

Details:
We followed the procedure on page 215 of P1300006.  The main idea is to use the crystal temperature at which red and green co-resonante as a reference for crystal position, since both the position of the cavity axis along the wedge and the crystal temperature set the co-resonance.  There are a few caveats to the way we did this:

These caveats can explain why this measurement has large scatter.  We did this measurement with about 6mW of green injected into the OPO, and we saw our best nonlinear gain between 15-20 for this input power.  We can do a better measurement of the OPO threshold by changing the input power at a single crystal position, but this data gives us an estimate of the threshold as between 10 and 11 mW without correcting for our mode mismatch, and between 7.4-8mW correcting for the 74% mode matching from alog 40594. 

The LLO OPO was expected to have a threshold of 45mW with an input coupler that is 87.5% reflective for 532nm, while our input coupler has a reflectivity of 98%.  The threshold power is proportional to the green cavity decay rate, so our expected threshold for this OPO is 7mW (45mW*(1-sqrt(0.875)/(1-sqrt(0.98)). 

Terry and I left the crystal at the position where we see co-resonance for 33.75C, because we think this is about at the peak of the phase matching curve.  This procedure will have to be repeated once HAM6 is pumped down.

 

Images attached to this report
Non-image files attached to this report
Comments related to this report
terry.mcrae@LIGO.ORG - 14:01, Wednesday 28 March 2018 (41183)

Sheila, Daniel, Nutsinee, Terry

The picomotor drive for the Crystal translation stage (the same drive is also separately used to shift the lens to mode-match the squeezer beam to the IFO) requires an in-house adaptor cable (attached PicomotorBreakoutCable.pdf and see D1700405) to go from the picomotor driver (attached PI_Datasheet_E-870_20150821.pdf) to the Squeezer interface chassis D1700185

Note the cable has two outputs, one for the crystal translation and one for the lens, and they cannot be used simultaneously. When required the particular drive cable for the crystal or lens is plugged into the "Translation Stage Picomotor Inputs" input of D1700185. The driver software is run from a personal computer.

Further documentation on the picomoter stage is here (D1500086).

Non-image files attached to this comment
Displaying report 1-1 of 1.