In trying to improve the mode matching into the pre-modecleaner, the alignment through the mode matching
lenses was redone because of a mechanical interference issue with already installed mounts. With the
lenses in their calculated positions, the beam was still too big for the pre-modecleaner and the visibility
was ~40%. More beam propagation measurements were taken to check the results from the previous week.
Re-doing the fit and allowing large variations in the amplifier output waist and location, resulted in new
positions for the lenses. That improved the pre-modecleaner visibility to ~75%.
At the moment, both the pre-modecleaner and reference cavity are locked but are by no means tweaked up.
I had hoped to be at this stage with less than 10 hours on the amplifier's clock but the alignment and
beam size checks took longer than expected.
Volker Quetschke / Peter