Displaying report 1-1 of 1.
Reports until 15:59, Thursday 30 June 2016
H1 SUS
betsy.weaver@LIGO.ORG - posted 15:59, Thursday 30 June 2016 - last comment - 14:46, Wednesday 04 April 2018(28090)
ETM15 Ears bonded

Between yesterday and today, Gerardo and I successfully bonded the ears to core optic mass ETM15, destined for LLO for post-O1 eventual replacement. 

Ear s/n 195 bonded to S3 flat of mass June 29, 2016

Ear s/n 196 bonded to S4 flat of mass June 30, 2016

The first ear (s/n 195) has a series of bubbles along it's edge, but the surface area of these do not add up to more than the allowed amount as per E1000277, so it passes, see pix attached.

The second ear (s/n 196) has no bubbles and is nice full bond. 

Images attached to this report
Comments related to this report
betsy.weaver@LIGO.ORG - 11:54, Friday 01 July 2016 (28118)

Inspection of the bonds this morning showed no change to the nice S4/Ear surface bond.

The small edge bubbles along the S3 Ear have migrated a bit (not worse, just moved a bit).  Therefore, we decided to rewet the edge with solution to see if it would wick in and close some of the bubbles.  This worked to close one of the larger bubbles, but one bubble continued to morph so we'll reinspect Monday.  Note, none of the sizes of the bubbles got bigger, so the surface area of the bubble total remains the same and below spec.

Images attached to this comment
betsy.weaver@LIGO.ORG - 10:02, Tuesday 05 July 2016 (28148)

Attached is a picture of what the bubbles in the ETM15 S3 Ear bond look like today (recall, it was bonded last Wed so it has now been over 5 days since initial bond).  Apparently over the weekend, the bubbles ran inbound a bit further.  I estimate the surface are of the bubbles to be near 20mm sq., still under the 50mm sq. spec. E1000278, although it isn't pretty.  I measured the longest streaks to be almost 3mm long, with a width of 0.4mm.  There are no bubbles inward from the edge near the center - anything shown in the picture is a camera reflection.

 

The S4 Ear bond looks unchanged from inspection last week, very good bond, very little bubbles.

Images attached to this comment
betsy.weaver@LIGO.ORG - 14:46, Wednesday 04 April 2018 (41291)

Pictures of this ear bond taken by Danny at LLO just before use at LLO, indicate that the bubbles have not changed shape or size since the picture attached July 2016.  Phew, said both Gerardo and I.

Images attached to this comment