J. Kissel
I've taken standard close-out transfer functions for H1 SUS TMSY. It's free and clear of rubbing, and results looking strikingly similar to before. OK for pump down!
Raw data is committed here:
/ligo/svncommon/SusSVN/sus/trunk/TMTS/H1/TMSY/SAGM1/Data
2018-04-04_2209_H1SUSTMSY_M1_WhiteNoise_L_0p01to50Hz.xml
2018-04-04_2209_H1SUSTMSY_M1_WhiteNoise_P_0p01to50Hz.xml
2018-04-04_2209_H1SUSTMSY_M1_WhiteNoise_R_0p01to50Hz.xml
2018-04-04_2209_H1SUSTMSY_M1_WhiteNoise_T_0p01to50Hz.xml
2018-04-04_2209_H1SUSTMSY_M1_WhiteNoise_V_0p01to50Hz.xml