Reports until 17:06, Wednesday 04 April 2018
LHO VE
chandra.romel@LIGO.ORG - posted 17:06, Wednesday 04 April 2018 - last comment - 19:12, Thursday 05 April 2018(41296)
Vacuum summary for the day

Gerardo, Mark D. Tyler, Chandra

Corner

EY

We will start roughing down EY tomorrow.

Comments related to this report
chandra.romel@LIGO.ORG - 17:22, Wednesday 04 April 2018 (41297)

Checked on CP4 at MY.

gerardo.moreno@LIGO.ORG - 17:49, Wednesday 04 April 2018 (41299)VE

We have an aux cart running at Y-End right next to the BRS, it will be there until pressure at the BSC10 annulus system is kept at nominal by its ion pump.

Images attached to this comment
jeffrey.kissel@LIGO.ORG - 08:00, Thursday 05 April 2018 (41301)CDS, ISC, SYS
Prior to corner station vent, high voltage has been turned off (as per items mentioned in M1300464).
    - ESD driver
    - Fast shutter de-energized
    - OMC's PZT actuator drive
    - Picomotors
ITM Ring heater's measured power is zero (power is off at chassis), and I've changed the requested power to zero via MEDM.
kyle.ryan@LIGO.ORG - 19:12, Thursday 05 April 2018 (41318)
The net pump speed of this diaphragm backing pump is "much" less than what we are used to. I used this portable pump station for the first time the other day to evacuate 4 lengths of 1.5" flex line + the volume between IP6's closed GV and the zero-length reducer and it took "minutes" not seconds.