Adjust the mirror mount before the rotation stage and the one at the base of the IO periscope to partially bring the beam back to the input modecleaner. One can certainly see flashes as MC2 swings. The knobs were moved less than 1/16 of a turn in both axes. I limited the power to the input modecleaner to 5 W by using the half waveplate and polariser combination at the output of the pre-modecleaner. The output of the photodetector on the table that monitors the power going into HAM1 did not change appreciably from ~60 mV. Since the new EOM is not resonant at 24.1 MHz and because of some changes to the RF signals, the phase and sign of the error for the input modecleaner may need some adjustment.