Displaying report 1-1 of 1.
Reports until 17:10, Tuesday 11 September 2012
H1 SUS
jeffrey.bartlett@LIGO.ORG - posted 17:10, Tuesday 11 September 2012 (4159)
Check MC2 for rubbing
   This afternoon Andres and I checked MC2 for rubbing. We found the two left side (under BOSEM T1) Lower Blade Stops almost touching their respective blades. We backed both stops off to closer to the specified 0.75mm spacing. We also found the upper left EQ stop on the Bottom mass was lightly touching the glass, which we backed off. We checked the remaining EQ stops and adjusted as necessary. MC2 should be ready to continue testing.    