jeffrey.kissel@LIGO.ORG - posted 11:53, Thursday 26 April 2018 (41687)
H1 SUS OMC TFs Free of Rubbing; OSEM Sensor Spectra Looks Roughly as Expected; OK for HAM6 Doors
J. Kissel

Grabbed a few closeout measurements of H1 SUS OMC. All clear; good to go for door closure.

Transfer functions show that the dynamics are virtually identical to what they were before. See 
    - Individual measurement 2018-04-25_2324_H1SUSOMC_M1_ALL_TFs.pdf 
    - Comparison with others allomcss_2018-04-25_2324_H1SUSOMC_M1_Phase3a_ALL_ZOOMED_TFs.pdf 

High Frequency OSEM sensor noise ASD,
    - 2018-04-26_1831_H1SUSOMC_M1_OSEM_Noise_ASDs.png
has some spikes, but nothing egregious like a grounded BOSEM (e.g. LHO aLOG 40787). If I had to complain, I would complain about the T1 OSEM being a little higher in noise starting around 500 Hz. But I don't have to complain, so I won't. #WORKSFORME

TF Data Templates 
    2018-04-25_2324_H1SUSOMC_M1_WhiteNoise_L_0p02to50Hz.xml
    2018-04-25_2324_H1SUSOMC_M1_WhiteNoise_P_0p02to50Hz.xml
    2018-04-25_2324_H1SUSOMC_M1_WhiteNoise_R_0p02to50Hz.xml
    2018-04-25_2324_H1SUSOMC_M1_WhiteNoise_T_0p02to50Hz.xml
    2018-04-25_2324_H1SUSOMC_M1_WhiteNoise_V_0p02to50Hz.xml
    2018-04-25_2324_H1SUSOMC_M1_WhiteNoise_Y_0p02to50Hz.xml

ASD Template 
    2018-04-26_1831_H1SUSOMC_M1_OSEM_ASDs.xml