Displaying report 1-1 of 1.
Reports until 15:35, Wednesday 02 May 2018
H1 SEI (SEI)
patrick.thomas@LIGO.ORG - posted 15:35, Wednesday 02 May 2018 - last comment - 16:27, Wednesday 02 May 2018(41803)
H1 ISI CPS Noise Spectra Check - Weekly
FAMIS 6948

BS_ST1_CPSINF_H1_I, BS_ST2_CPSINF_H1_I and BS_ST2_CPSINF_V1_I appear high.
ITMX_ST1_CPSINF_H2, ITMX_ST1_CPSINF_V2, ITMX_ST2_CPSINF_H2 and ITMX_ST2_CPSINF_V2 appear very low.

Scripts reported:

BS_ST1_CPSINF_H1_I high freq noise is high!
ETMX_ST2_CPSINF_H1 high freq noise is high!
ETMX_ST2_CPSINF_H2 high freq noise is high!
ETMX_ST2_CPSINF_H3 high freq noise is high!
ETMX_ST2_CPSINF_V1 high freq noise is high!
ETMX_ST2_CPSINF_V2 high freq noise is high!
ETMX_ST2_CPSINF_V3 high freq noise is high!
Images attached to this report
Comments related to this report
hugh.radkins@LIGO.ORG - 16:27, Wednesday 02 May 2018 (41808)

Thanks Patrick. I agree with your assessments.  The ITMX issue appears to be a shutdown of the satellite box, I suspect, related to the door operations as it went zero late morning yesterday.  I don't suspect a failure at this point but maybe the power supply has unplugged--it should be screwed in but if not then it could have been disrupted.  I'll open a ticket and investigate.  FRS 10551.