hugh.radkins@LIGO.ORG - posted 09:11, Wednesday 16 May 2018 (42010)
WHAM5 ISI Corner3 FF L4C Pod Pressure Electronics Suspected of Failing---Solved

Re aLog 41668, FRS 10499

Saw the Pod Pressure at the ADC rail.  Been this way since April 12.  Investigation found unsecured (screws likely not at all engaged) field cable going into the Interface chassis.  Better seated connector and problem solved.