Displaying report 1-1 of 1.
Reports until 09:16, Wednesday 23 May 2018
H1 AOS
aidan.brooks@LIGO.ORG - posted 09:16, Wednesday 23 May 2018 (42139)
ETM HWS source replacement - status update day 2 - return beam from ETM detected. Poor beam quality

[Aidan, Marie, Dan B, Alexei, Keita]

Yesterday morning we realigned the ALS beam into the vacuum system, confirmed it was centered on the TMS QPDs, confirmed it was coming out again and was centered on the in-air WFS and confirmed that it was getting back to the ALS length PD. The two irises installed after the PBS were centered on this ALS beam.

We then aligned the LED beam through the PBS and through the two irises. By carefully fine-tuning this alignment as best we could by eye, we found that there was a return beam coming back to the Hartmann sensor.

The return beam is quite sensitive to alignment. There is a narrow window of adjustment of the pitch and yaw on the final two HWS mirrors. Small changes (fractions of a turn) can make the beam disappear entirely. We don't have picomotors on these actuators so we can't easily quantify what "small" means here.

We were able to get images of the return beam on the Hartmann sensor camera. We started with the Hartmann plate in place but eventually took it off and saw the intensity distribution of the return beam. It looks quite mottled and smaller than expected. 

We spent the remainder of the day trying to improve the spatial quality of the return beam. It looks like it might be clipped in multiple places or is experiencing some diffraction off particulates on optics - we're still investigating. We experimented with trying to change the divergence of the beam coming from the fiber launcher as well as removing the iris that was in front to get more of the beam through.

Due to concerns about clipping, we temporarily removed the PBS so that the HWS beam didn't have to contend with that small aperture.

The goals today are to:

 

Images attached to this report