Reports until 10:09, Tuesday 29 May 2018
H1 PSL (PSL)
peter.king@LIGO.ORG - posted 10:09, Tuesday 29 May 2018 (42205)
power noise
Following on from measurements taken  last Wednesday  more
measurements of the power noise were taken this morning.  Measurements were after the neoVAN power amplifier, before and after the
pre-modecleaner.  The broad peaks at ~18.5 kHz and multiples is present on the output of the amplifier.  It is possible to that
this might be due to an incorrectly set crossover in the FSS but I didn't think to check this when I was in the enclosure.

    The photodiode used for monitoring the output of the amplifier has a broad peak at ~5 kHz, which should not be there.

    The power noise measurements were taken with the ISS on and off.  Clearly the ISS is suppressing the power noise but there are
a number of 60 Hz related peaks in the spectrum what need to be eliminated.
Images attached to this report