Displaying report 1-1 of 1.
Reports until 09:33, Wednesday 30 May 2018
H1 SYS
daniel.brown@LIGO.ORG - posted 09:33, Wednesday 30 May 2018 (42225)
Testing AS port protection

Dan, Jenne

We tested the AS port protection by setting the fast shutter threshold low. The shutter triggered and the sum on AS_A/AS_B dropped. The AS port protection MEDM correctly signaled that the shutter closed and the screen elements went red. We reopened the shutter and light is coming through but the fast shutter screen says there is an "error" but doesn't provide any more information, screenshot attached.

Images attached to this report
Displaying report 1-1 of 1.