Displaying report 1-1 of 1.
Reports until 14:40, Friday 01 June 2018
H1 SUS
betsy.weaver@LIGO.ORG - posted 14:40, Friday 01 June 2018 - last comment - 10:06, Tuesday 05 June 2018(42285)
ETMX ETM16 Test Mass HR Surface features observed during closeout

While inspecting the HR surface during (and just) after the FirstContact pull on the ETMX HR Surface yesterday, Travis and I saw a few points within the central 6" of the optic.  Pictures are attached.  One feature looks more like a small sleek or scratch rather than a dot.  Upon review of the "map" we made roughly a year ago when the optic was in the bonding lab (also attached), we see that many of these were there during that inspection.  In conference with GariLynn this morning, she confirmed my decision to move to chamber closeout and the door has just been installed on the chamber.  We will digest these pictures and compare to previously obtained coating maps and metrology data.

Note, keep in mind that it is very difficult to say much about the optic surface quality based on these photos.  It is difficult to state whether such observed features will be a problem, whether they are in the coating or on top of the coating, etc.  TBC...

 

Images attached to this report
Comments related to this report
garilynn.billingsley@LIGO.ORG - 16:22, Friday 01 June 2018 (42295)COC
The scratch is found in the original polishing defect report at C1205029
Scatter measured with an integrating sphere shows several high points in the center 50 mm diameter, measured scatter is 7.7 ppm at E1500336
Thanks for the great pics!
garilynn.billingsley@LIGO.ORG - 10:06, Tuesday 05 June 2018 (42327)
Further analysis at https://dcc.ligo.org/T1800255  No new conclusions.
Displaying report 1-1 of 1.