Laser Status:
SysStat is good
Front End Power is -0.005834W (should be around 30 W)----->This has been 0W since March 19, 2018, due to PSL work...should this check be removed/updated?
HPO Output Power is 0.926W
Front End Watch is GREEN
HPO Watch is RED
PMC:
It has been locked 6 days, 18 hr 55 minutes (should be days/weeks)
Reflected power = 18.51Watts
Transmitted power = 42.82Watts
PowerSum = 61.33Watts.
FSS:
It has been locked for 0 days 15 hr and 37 min (should be days/weeks)
TPD[V] = 2.8V (min 0.9V)
ISS:
The diffracted power is around 0.87% (should be 3-5%)
Last saturation event was 0 days 16 hours and 53 minutes ago (should be days/weeks)
Possible Issues:
Front End Power is Low (see highlighted note above)
PMC reflected power is high
ISS diffracted power is Low
LRA out of range, see SYSSTAT.adl