chandra.romel@LIGO.ORG - posted 15:15, Monday 02 July 2018 - last comment - 13:27, Monday 09 July 2018(42737)
EX RGA scans

Attached are two RGA scans:

Later in the week we will collect more scans to see how the water is falling.

Total pressure at EX before valving in CP was 5e-7 Torr and is now 5e-8 Torr, as expected, as a result of a full order of magnitude decrease in water partial pressure.

Note that the short CPs allow 30% of water molecules to escape past the trap and into BT; long CPs in corner allow only 3% transmission.

EX is currently isolated from BT with a soft closure of GV19. We typically don't operate in this state (BT valve soft closed with a VEA turbo valved in) in case the turbo pump fails and back streams into main volume, possibly seeping past GV o-ring.

Images attached to this report
Non-image files attached to this report
Comments related to this report
chandra.romel@LIGO.ORG - 14:21, Tuesday 03 July 2018 (42755)

Tuesday 2pm RGA scan 26 hours after valving CP into EX volume.

Non-image files attached to this comment
chandra.romel@LIGO.ORG - 09:56, Thursday 05 July 2018 (42766)

RGA scan after 67 hours with CP8 valved in (plus 2000 l/s turbo).

Non-image files attached to this comment
chandra.romel@LIGO.ORG - 15:41, Friday 06 July 2018 (42793)

EX RGA scan after 99 hours with CP8 valved in. This scan was taken one day after we temporarily (8-hr day shift) opened to x-arm.

Non-image files attached to this comment
chandra.romel@LIGO.ORG - 13:27, Monday 09 July 2018 (42817)

Correction on short CP H2O transmission. Value is 17.2% (not 30%), per C960963. Long pumps (in corner station) are 2.6% each.