Displaying report 1-1 of 1.
Reports until 14:46, Monday 09 July 2018
H1 PSL (PSL)
jeffrey.bartlett@LIGO.ORG - posted 14:46, Monday 09 July 2018 (42821)
PSL Weekly Report FAMIS #7498)
Laser Status:
SysStat is good
Front End Power is -0.002217W (should be around 30 W)
HPO Output Power is 0.4705W
Front End Watch is RED
HPO Watch is RED

PMC:
It has been locked 3 days, 6 hr 19 minutes (should be days/weeks)
Reflected power = 28.8Watts
Transmitted power = 33.62Watts
PowerSum = 62.42Watts.

FSS:
It has been locked for 0 days 0 hr and 2 min (should be days/weeks)
TPD[V] = 1.973V (min 0.9V)

ISS:
The diffracted power is around 0.11% (should be 3-5%)
Last saturation event was 0 days 0 hours and 0 minutes ago (should be days/weeks)

Possible Issues:
Front End Power is Low
PMC reflected power is high
ISS diffracted power is Low
LRA out of range, see SYSSTAT.adl
Displaying report 1-1 of 1.