H1 TCS
stefan.ballmer@LIGO.ORG - posted 21:24, Monday 09 July 2018 - last comment - 10:19, Wednesday 11 July 2018(42831)
ITMY Hartmann work

Daniel, Thomas, Stefan

- We added a button to the TCS-HWS medm screen that plots the contour difference between a reference time and the current time. For now the reference time is taken from /opt/rtcds/userapps/release/tcs/common/medm/referencetimeITMY.txt

- We really want a streaming version of this working, but the current streaming scripts are inoperational (and some of them interfere with the EPICS logging)

- With that we did a quick shot with the CO2 laser - it didn't look good... (plot 1) It almost looks like CO2 heating fringes... we will have to go back to this.

- We next wanted to verify that none of this is the fault of the HWS, so we did a ring heater run. That one looked reasonably well (plot 2), both in circularity and centering (note that the centering is sensitive to SR3 alignment).

- The optics are now cooling down over night - we will look at the CO2 heating again tomorrow morning.

Images attached to this report
Comments related to this report
thomas.vo@LIGO.ORG - 10:40, Tuesday 10 July 2018 (42835)

We left ITMY ring heaters and CO2 lasers off last night and attached is the cool down between Jul 10 2018 5:00:00 UTC and Jul 10 2018 16:00:00 UTC.

It seems like the ITMY HWS beam is still pretty well centered on on the test mass based off the ring heater cool down pattern.

Images attached to this comment
thomas.vo@LIGO.ORG - 10:19, Wednesday 11 July 2018 (42858)

To verify the calibration of the HWS with the ring heaters, we tried to match the COMSOL model of the transient heating here with the readout of the HWS sensing the RH heating (picture attached).

From Aidan/Stefan's remarks about the linear part of the RH absorption, the lensing effect can be approximated by

 dS/dt = 6.75E-9 diopters per second per Watt

Using this equation we estimate that the Comsol model agrees with the HWS up to about 20% which is roughly consistent with previous tests.

Method Slope of thermal lensing in linear regime
COMSOL 0.10 microdiopters/sec
HWS Spherical Power 0.08 microdiopters/sec
Images attached to this comment